Technology ID
TAB-2824

Improved Acoustic Plethysmograph System for Noninvasive Measurement of Pulmonary Function

E-Numbers
E-290-2013-0
Lead Inventor
Reynolds, Jeffrey (CDC)
Co-Inventors
Frazer, David (CDC)
Applications
Vaccines­­­
Therapeutics
Software / Apps
Research Materials
Occupational Safety and Health
Non-Medical Devices
Medical Devices
Diagnostics
Consumer Products
Therapeutic Areas
Pulmonology
Ophthalmology
Oncology
Infectious Disease
Immunology
Endocrinology
Dental
Cardiology
Development Stages
Pre-Clinical (in vitro)
Development Status
  • In vivo data available (animal)
  • In situ data available (on-site)
  • Prototype
    Research Products
    Research Equipment
    Computational models/software
    Lead IC
    CDC
    ICs
    CDC
    CDC researchers have developed a novel acoustic whole body plethysmograph (AWBP) that allows measurement of tidal volume in lab animals, independent of gas compression in the lung. This system provides particular advantages over the traditional whole body plethysmograph (WBP) when measuring model animals with increased gas compression due to increased airway resistance or increased acceleration in the breathing pattern.

    Measurement of tidal volume in conscious, unrestrained mice has traditionally been performed using WBP. An animal is placed in a chamber where pressure changes due to respiration are observed, which are then related to tidal volume. Although the effects of gas compression on the WBP signal of normal mice may be negligible, gas compression in mice with altered breathing pattern and/or increased airway resistance can produce significant errors in the measurement of tidal volume. A major advantage of this novel AWBP technology is the ability to measure the tidal volume signal independent of gas compression. This is particularly useful when measuring mice that have been exposed to a respiratory irritant or toxin that increases airway resistance or the frequency content of the breathing pattern. Further, with slight hardware and software modifications the system may also be used to measure specific airway resistances.
    Commercial Applications
    • Measurement of tidal volume, respiratory rate, and other breath rate parameters of laboratory animals
    • Research and animal modeling tool for in vivo pulmonary investigations of therapeutics efficacy and studying certain infectious diseases
    • Particularly useful when measuring animals exposed to a respiratory irritant or toxin that increases either airway resistance or the frequency
    • Noninvasive measurement of specific airway resistance and restrictions in an unrestrained animal
    Competitive Advantages
    • System allows fast, efficient, noninvasive measurement of lab animal pulmonary function for numerous inhalation, toxicology research studies
    • Measures tidal volume signal independent of gas compression, increasing accuracy when gathering data from animals with altered breathing patterns and/or elevated airway resistance
    • System operates at a fixed frequency, and automatically tracks the change in sound amplitude as a function of time due to change in lung volume
    Licensing Contact:
    Mitzelfelt, Jeremiah
    jeremiah.mitzelfelt@nih.gov